Multimeter metrology

Kragen Javier Sitaker, 02020-11-01 (updated 02020-11-27) (23 minutes)

A standard TL431 voltage reference is only accurate to ±0.5% at best, more often ±2%. TI’s REF5050 is accurate to ±0.1% or ±0.05% in the “high-grade” version, with 3 ppm/K temperature drift and 50 ppm/1000 hours, but those cost US$5–10. The internal voltage reference in the STMF103C8 used in many Blue Pill boards is specified as 1.16–1.26V over the -40° to 105° range, an error of about ±4.2%, with a temperature coefficient of 100 ppm/K. The datasheet for the CKS32 in the Blue Pill I got says the same thing in §5.3.4 内置的参照电压: its VREFINT (内置参照电压) goes from 1.16V (最小值) to 1.26V (最大值).

This is pretty shitty fucking accuracy. 4.2% is 42000 ppm. For measuring temperature with a tungsten wire near room temperature, a ±4.2% error is a ±10° error, and it gets larger at higher temperatures. Using a standard TL431 ripped out of some scrapped power supply we can cut that error to 19000 ppm or so, but that’s still nothing to write home about. How the fuck are you supposed to build a fucking voltmeter? Or anything that depends on voltage for accuracy?

By contrast, the quartz crystal on the Blue Pill is 8 MHz with, probably, an error of about 10 ppm, common for watch crystals. The damn chip can measure voltages to 1.5 digits of accuracy and time to 5 digits of accuracy, 4200 times better. (And it can count electrical pulses to a lot more accuracy than that, to the point that the concept of tolerance stops being meaningful.) We can get down to probably 1 ppm timing error if we can measure the temperature to compensate, 0.1 ppm if we can control it. (And nowadays we can reference to GPS.)

Like, something similar to the open-source Transistortester AVR by Karl-Heinz Kübbeler and Markus Frejek, now commonly known as the “M328 Transistor Tester” (Instructables) (English manual) or “LCR TC1 ESR meter”, which sells for US$38 here in Argentina or US$13 in the US. But with better accuracy, precision, and repeatability, and ideally without any bought components.

The 0.1%-precision LM4040CIM3X-20-NOPB voltage reference costs US$1.67 from Mouser.

How good are multimeters usually?

Some random YouChube video from TheSignalPath (#175 (ⅱ)) found about 25 ppm of voltage difference between his not-recently-calibrated Fluke 744 Documenting Process Calibrator and his not-recently-calibrated 8-digit Keithley DMM 7510, so tens of ppm is achievable in ordinary electronics labs, but expensive; he apologized for the lack of calibration, so apparently he was expecting better. His resistance error was worse, 4600 ppm at 100Ω, 400 ppm at 1kΩ, 20 ppm at 10kΩ; this pattern makes me suspect that this error was due to not using a 4-wire measurement. His current error was about 250 ppm. Testing the 744 against its own meters he instead got 400 ppm emf error and 500 ppm resistance error; he neglected to thus test its current measurement capability.

I tested a shitty digital multimeter from 2016, US$6 from the hardware store, against another similar meter to get an idea of how bad they are, and also how aggressive they are. It seems like the old multimeter’s diode-testing range uses up to 1.4 mA at up to 2.72 volts. Their diode test readings differ by about 2%, resistance by about .03%, voltage by about .3%. pretty impressive metrology for US$6. These are lower bounds on their actual errors, but they’re unlikely to be conservative by more than an order of magnitude or so.

Canceling out voltage errors in measurements

Ratiometric voltage measurements with the dual ADCs ought to be a lot more precise, and probably limited only by the 12-bit bit depth and whatever the noise is. So, for example, if we run an exactly known current through a tungsten lightbulb and measure the voltage, we’re subject to that shitty ±4.2% precision; but if we put some unknown voltage across a series combination of the lightbulb and an exactly known resistance, we can get a much higher-precision measurement.

Resistance standards

How do we get an even approximately known resistance, though? Typical resistors used to be ±20%, now they’re ±1%, but nothing close to 1 ppm. On MercadoLibre we can buy a temperature standard resistance made of Weston’s manganin, intended for converting precise voltage measurements to current measurements or vice versa, for US$20. At 25° manganin’s temperature coefficient of resistance crosses zero, having fallen from 6 ppm/K at 12° on its way to -42 ppm/K at 100° and -52 ppm/K at 250°, before crossing zero again at 475°. This suggests that a ±10° error of temperature measurement around 25° would give you about a ±5 ppm error of resistance. However, these resistors are not so precisely made, having a ±0.5% error, 5000 ppm; one wonders if they are correctly annealed or even truly made of manganin. Better precision calibration resistors are available.

A better option might be a known capacitance; you can easily measure the RC time constant as long as your voltage reference isn’t too noisy over the short time involved. C = εA/d, so it’s straightforward to construct a capacitor with a known capacitance; however, if we want its capacitance to be known to within, say, 10 ppm, we need less than 10 ppm error on all of ε, A, and d. A is relatively easy: a 1 m x 1 m square of foil is a square meter to within 10 ppm if its width and height are each accurate to within 7 ppm (i.e., an average of 7 μm) and it’s square to within cos⁻¹(1-10ppm), which is about 15 minutes of arc. Getting d accurate to within 10 ppm is more difficult; if the dielectric is intended to be 100 microns, its average thickness must be correct to within 1 nm. But getting ε accurate to within 10 ppm is feasible in only one way: vacuum. Even the permittivity of air is 590 ppm higher than vacuum, and it varies according to pressure and temperature. Common solid dielectrics are hopeless; WP gives polypropylene’s relative permittivity as “2.2–2.36”, i.e. ±35000 ppm. A vacuum capacitor of these dimensions would have a capacitance of 88.5419 nF.

Oversampling

Every time we quadruple the measurement time, we add another bit of precision, because the variances of Gaussian noise combine additively. So the sum of four measurements has four times the variance, thus twice the standard deviation, as a single measurement, and their average thus has half the standard deviation. The Blue Pill’s ADCs run at 1Msps, so if we take 1048576 samples, 1.05 seconds’ worth, we can get a 22-bit-precision reading, which has a quantization error of ±0.125 ppm. Probably averaging 65536 or 32768 samples is a better tradeoff, giving you more like 1 ppm error in exchange for much faster data acquisition.

Hamer’s monograph on Clark, Daniell, and mostly Weston cells

In despair, I turned to a shitty scan of NBS Monograph 84, from 1965-12-15, titled, “Standard Cells: Their Construction, Maintenance, and Characteristics”, by one Walter J. Hamer. He explains to some extent the construction of the Daniell cell, the Clark cell, and the Weston cell, which served as laboratory references from 1836 until 1990. A Daniell cell, using copper, zinc, and their sulfates, is relatively straightforward to construct, but drifts rapidly over time owing to the mixing of the sulfates, depends strongly on the purity of the copper and the zinc, and has a horrific temperature coefficient. The Weston cell was the first to have a roughly zero temperature coefficient, which it achieves by the use of cadmium salts, which unfortunately are rather difficult for me to procure.

Hamer also goes into the history of other metrological standards for voltage (a word he disdains to use, preferring “emf”) and other electromagnetic units. He points out that in theory you can use an absolute electrometer as a standard, using the permittivity of free space and precise measures of current, but that this gives an error on the order of 100 ppm. (He doesn’t go into detail, but it turns out this is a matter of charging a known capacitance, constructed as above, to a known voltage or charge and measuring the resulting force; one version is an electrostatic balance where the electrostatic repulsion between two plates is countered by a sufficiently precise weight to return them to their original position.)

I suspect that you could construct a microscopic absolute electrometer that would give you much more precise readings; all the absolute electrometers I can find in the literature were on the order of 100 mm in diameter with on the order of 10 mm plate separation, and so involve weights on the order of 10 N but electrical forces on the order of 10 mN when operated at a few kV, and are additionally tricky to operate under vacuum. If you were to scale it down by a factor of 10,000, you would have capacitors with 10-μm-diameter plates separated by 1 μm, weighing on the order of 1 ng; with those precise dimensions it would be 0.88541878 femtofarads. But what would the force be?

In a parallel-plate capacitor, C = εA/d = q/V = 2E/V², so the capacitance varies directly with the plate area, inversely with the plate spacing, inversely with the voltage at a fixed charge, directly with the energy at a fixed voltage, and inversely with the squared voltage at a fixed energy. If you were to start with the plates in contact (but miraculously not discharging; maybe they’re triboelectric insulators like packing tape and its adhesive) the capacitance would be infinite, so the voltage would be zero. If you then pull the plates apart to some distance, you have to add some energy as the capacitance starts to fall and the voltage correspondingly to rise, proportional to the distance.

Now, if we measure dE/dd at some distance then we get the force between the plates there. E = ½CV², and as we pull the plates apart at a fixed charge the capacitance drops (proportional to distance) and the voltage rises (also proportional to distance): q/V = εA/d, so V = qd/εA, so V² = q²d²/ε²A², and E = ½CV² = ½εAq²d²/ε²A²d = ½dq²/εA. So that means the force between the plates is constant: dE/dd = ½q²/εA regardless of the distance (as long as the parallel-plate approximation is valid). So for example if we charge this cap to 1 kV it would have 0.88541878 picocoulombs on it, producing a force of 0.88541878 millinewtons.

Actually you can’t charge it up that far because after 20–40 MV/m you get field emission across the vacuum which limits you to 20–40 V/μm. Say you charge it to 10 V instead. Now it’s only 88.541878 nanonewtons.

That doesn’t sound like a whole lot, but it’s 8000 times larger than its weight instead of a thousand times less, so you could easily make it overwhelmingly the largest force on the capacitor. It’s a lot less than an atmosphere, though.

As a standard of resistance, he describes the construction of an air-core inductor of known dimensions and thus computable inductance, and the measurement of its E–I relationship at different frequencies to obtain a precise standard for the ohm; but he describes the Wenner method, which seems to be some kind of differential measurement that I don’t fully understand, getting a ±5 ppm precision. He also mentions using the rotation of a magnet in a coil, or a coil in the earth’s magnetic field. (The quantized Hall effect is the modern absolute standard since 1990.)

He says that using computable capacitors (like my microscopic thought experiment above) to check the ohm measure is “less involved” and “may be used [text lost] an annual basis” to check resistance standards against absolute units in preference to the inductive approach. (He also says they normally used 1-pF computable capacitors rather than the 88000-pF jobby I used above for calculations.) “Thompson-Lampard theorem” seems to be the key term here.

Given the possibility of measuring a computable inductor or computable capacitor with a microcontroller, I’d think that it would be easier to do the measurements in the time domain rather than the frequency domain.

Note that a precise standard of resistance, or a precise standard of both time and either inductance or capacitance, allows you immediately to convert between precise current and precise voltage. Time is, as I said, easy enough now.

Hamer then explains that an absolute standard of current is available in the form of a current balance, in which you measure the electromagnetic repulsion or attraction between two conductors against a standard weight, and of course this is the standard definition of the ampere; he says this is about ±6 ppm. The method he describes requires some kind of measurement over time, but I think current balances are more commonly used in a quasistatic fashion, in which after adding a mass, the current is increased until repulsion returns the mass to its original position, thus requiring only measurement of length and mass to calibrate. And I think that in fact they are commonly used nowadays for measuring masses in terms of known currents rather than vice versa.

Force balances have the stunning metrological advantage that the difference in force produces an acceleration, the acceleration integrated over time produces a velocity, and the velocity integrated over time produces a displacement. So even a very small force imbalance can produce an easily visible displacement, particularly if you use it to deflect a mirror reflecting a laser pointer across the room.

It’s feasible, though not an everyday occurrence, to get weights that are calibrated to within 1 ppm, though care must be taken to compensate for local gravitational fields and atmospheric pressure and humidity. Atmospheric pressure can diminish by as much as 14% in hurricanes, and of course varies by much more when you go up in elevation or underwater; a 100 g steel weight that occupies 12.6582 mℓ under some conditions is displacing about 15 mg of air, diminishing its apparent weight by 150 ppm. This can decrease by 2.5% from humidity or 14% from weather-related pressure variation, causing a deviation of some 20 ppm. Thermal expansion and contraction of the weights is not a concern; the 36 ppm/K volumetric expansion coefficient of steel means that the above-described weight will occupy 12.6628 mℓ at a temperature 10° higher, displacing about 5 μg more of air, lowering its apparent weight by 5 μg or 0.05 ppm.

Joe blocks, similarly, can measure lengths on the order of 100 mm down to a precision of 0.1 μm or so, though they expand and contract by 12 ppm/K, thus requiring 80 mK control to reach 1 ppm.

The Clark cell, Hamer says, was adopted in 1893 as 1.434 volts at 15°, though the modern value is 1.4328, and its temperature coefficient is -1.15mV/K. It uses a zinc or zinc amalgam anode, a mercury cathode, and a saturated aqueous electrolyte of zinc sulfate, plus mercurous sulfate paste as a depolarizer, to prevent hydrogen buildup on the plate (converting it instead into oil of vitriol; removing impurities in the mercurous sulfate lowered the voltage by another 300 μV). This would probably also be challenging for me to fabricate. We can estimate that a ±10° temperature error would result in a ±11.5mV error, about ±8000 ppm, which is why the Weston cell was adopted as the new standard in 1908.

The Weston cell has a temperature coefficient of about 41 ppm/K, substantially better than the 800 ppm/K given above for the Clark cell.

We can see why the NBS controlled the temperature of its Weston-cell room to within 1°, the temperature of its oil baths for the Weston cells to within 10 mK, and the oil baths’ temperature during measurements to within 1 mK. Such measures applied even to the Clark cell would have reduced its temperature-induced voltage error to some 0.8 ppm, at which point other errors would surely dominate.

Calibrating voltage by measuring power in a bomb calorimeter

It occurs to me that power might be a useful way to calibrate voltage given a known current or especially a known resistance, because if you can measure the power of a known resistance to within 1000 ppm, you know the voltage to within 1 ppm; moreover, ΔT = ∫P dt / Cth, so you get the same kind of integration effect as with force balances, but only to the first order, not the second order. Somewhat surprisingly, although specific heats vary with temperature, some are in fact known to the required precision; for example, according to definitions.units, the energy to raise a gram of water from 14.5° to 15.5° is 4.18580 J, the energy to raise it from 19.5° to 20.5° is 4.18190, and the energy to raise it from 0° to 100° averages 4.19002 per degree. So you could imagine, for example, heating 1000 kg (±1 kg) of very-well-insulated water by applying an unknown voltage to a known resistance (±1 ppm) from 0° to 100° (each ±10 mK) and measuring the time required (±1000 ppm), and thus getting a measurement of power to a precision of 1000 ppm, and thus measuring the voltage to 1 ppm. The imprecision of the specific-heat number only adds about 1.3 ppm to the power imprecision, and thus 1.7 parts per trillion to the voltage imprecision.

XXX hmm, maybe I screwed up that resistance calculation too?

XXX all of this stuff about calculating from power is based on a wrong logic step. 1000 ppm power error gives you 500 ppm voltage error (499.9 to be exact), not 1 ppm.

The main sources of imprecision in such an experiment would seem to be the insulation of the water, which would have to leak less than 0.4 joules during the experiment, and the original measurement of the resistance.

Computable inductors are easier to isolate from the environment

Which brings us back to the resistance-measurement problem. It occurs to me that it a computable inductor might be a more precise way to measure a resistance, particularly if it can be made very small in physical dimensions so that its magnetic dipole does not impinge on materials with a substantial permeability; wood’s, for example, is 0.43 ppm higher than the vacuum, a situation three orders of magnitude more promising than the corresponding situation with capacitors, and teflon’s is even closer (how much closer is not known).

Differential measurement of computable capacitors or inductors

A different tack might be to use some kind of differential measurement to precisely calibrate the dielectric constant or permeability of some material whose presence cannot be avoided but whose quantity can be varied. Operating the same air-gap capacitor at various air pressures, for example, might enable you to extrapolate its capacitance at hard vacuum, without having to actually produce 100-mPa vacuums.

Sensors

All these “sources of error” can equally well be seen as “observable variables”. The only difficulty is untangling them. If your capacitor’s resonant frequency (with a given coil) varies linearly with the air pressure, then by measuring that resonant frequency with 1 ppm accuracy, you can measure the air pressure with 1-ppm accuracy. If it also varies dramatically with the air’s moisture content, well, congratulations, you have a moisture sensor too, as long as you have some other way to sense air pressure that varies differently with humidity. (Inversely, ideally, or failing that, not at all.) Everything varies with temperature, but if some things vary more than others and you can keep it at the same temperature, you can measure the temperature. If you find something that varies a lot with temperature, maybe like the leakage current in a Schottky, you have a very precise temperature sensor, which you can use to cancel out temperature effects on other things, as long as you can characterize them. Force of gravity makes your watch crystal run faster when it’s sideways? Great, kid, you gotcherself a MEMS accelerometer that costs 25¢.

A photoelectric-effect voltage standard?

Can we use light of a precisely known color — a low-pressure sodium line, for example — to produce photoelectrons with a precisely known energy, and thus a photocell with a precisely known cutoff potential? Unfortunately the work function of the cathode material is subtracted from that energy, and this depends on the purity (and identity, and crystal grain orientation) of the surface material, and the temperature also affects the cutoff potential slightly. Typical variations between differently prepared samples of the same element are ±10%.

Still, I suspect this might be a feasible way to get some sort of voltage standard, even if not a very good one.

It might be possible to precisely measure aging-stable photoemission in air if you use a metal with a sufficiently conductive and nonhygroscopic oxide, such as zinc.

You also get photoelectric emission inside solid-state semiconductor junctions; that’s how photodiodes work. However, I suspect the cutoff voltage for this solid-state photocurrent may be fuzzier, just because of the messy nature of condensed matter.